Electro-Technical Calibration

Home → Services → Calibration & Research Division → Electro-Technical Calibration

Electrical measuring instruments have applications in all kinds of industries, from food and pharma to automotive, manufacturing, and the defence sector. Hence, the calibration of these electrical instruments plays a significant role in ensuring safety and quality in these industries. It is also mandated by regulation for certain industries to have their instruments and devices calibrated at periodic intervals. Calibration ensures that the measurement values given by these devices are true to the value and hence ensure smooth operations. Calibration also guarantees the traceability of the measurements.

Our laboratory is climate-controlled, with the temperature maintained between 25 ± 2 ̊C and the humidity between 45% and 65%. Our electro-technical calibration lab is equipped with state-of-the-art test benches for the lab standards. Master equipment with traceability to national or international standards is available for the calibration of instruments.

The following details the list of instruments, including their respective parameters, our calibration range, and standard specifications.

S.N.Source / MeasureInstruments NameParameter to be calibratedRangeStandard Used
1SourceMultifunction / Multiproduct Calibrator (Fluke – 5550, 9100, 5080, 5500, 5502, 5520, 5522, 5530) Transmile / Zeal 
Process Calibrator
1. AC Voltage
2. AC Current
3. DC Voltage
4. DC Current
5. Frequency
6. Resistance
7. Capacitance
8. Temperature Simulation (RTD & Thermocouple)
9. Scope
10. DC Power
11. AC Power
12. Power Factor / Phase Angle
1. 10 µV to 1000 V @ 50 Hz to 10 kHz
2. 10 µA to 20 A @50 Hz to 10 kHz, 30 A @ 50 Hz
3. 10 µV to 1000 V
4. 1 µA to 30 A
5. 100 mHz to 1 GHz
6. 1mΩ to 1 GΩ
7. 1 pF to 100 μF @ 1 kHz, 1 μF to 100 μF @ 100 Hz
8. Thermocouple (All Types):  -250 ℃ to 2300 ℃
RTD (All Types):  -200 ℃ to 800 ℃
9. Time Marker (Oscilloscope):  10 nsec to 5 sec
DC Amplitude (Oscilloscope):  1.25 mV to 100 V
AC Amplitude (Oscilloscope):  5 mV to 100 V
Bandwidth (Oscilloscope):  50 kHz to 1000 MHz
10. 10 mW to 20 kW
11.  30 mW to 20 kW
12. 0.01 to UPF @ 50 Hz
(1 phase – (30V to 480V), (0.01 A to 20 A)
1. 8.5 DMM (Fluke)
2. 6.5 DMM (Fluke)
3. AC Measurement Standard (Fluke
4. LCR Meter (Keysight)
5. Precision Power Analyzer (Yokogawa)
6. 3 Phase Reference Standard (ZERA)
7. Universal Frequency Counter (Keysight)
8. USB RF Power Sensor (Keysight)
2SourceAC/DC Power Supply
AC/DC High Current Source

1. AC Voltage
2. AC Current
3. DC Voltage
4. DC Current
1. 10 µV to 1000 V @ 50
2. 2. 10 µA to 20 A @50 Hz to 10 kHz, 20 A to 1000 A @ 50 Hz 3. 10 µV to 1000 V
4. 1 µA to 1000 A
1. 8.5 DMM (Fluke)
2. 6.5 DMM (Fluke) 3. High Current Shunt
3SourceHigh Voltage Source / Oil BDV TesterAC & DC High Voltage1 kV to 100 kV @ 50 Hz
1 kV to 100 kV
Capacitive Voltage Divider
4SourceFrequency Generator / Signal Generator 1. Frequency
2. RF Power
3. Time Period
1.100 mHz to 6 GHz
2. (100 kHz to 6 GHz):  -42 dBm (5 mVpp) to 21 dBm (7.09 Vpp)
3. 1 nsec to 1 sec
1. Universal Frequency Counter (Keysight)
2. USB RF Power Sensor (Keysight)
5SourceResistance Box / Standard Resitor
Capacitance Box / Standard Capacitor
Inductance Box / Stanadar Inductor 
1. Resitance (DC)
2. Resistance (@1 kHz)
3. Capacitance (@1 kHz) 3. Capacitance (@100 Hz)
4. Inductance (@1 kHz)
1. 100 µΩ to 20 GΩ
2. 1 Ω to 100 kΩ
3. 1 pF to 1 mF 4. 1 µF to 1 mF
5. 100 µH to 10 H
1. 8.5 DMM (Fluke)
2. 6.5 DMM (Fluke)
3. LCR meter
4. MFC (Fluke)
6aMeasure Precision Multimeter (8.5 / 7.5 / 6.5)
Digital Multimeter (up to 5.5 digit)
Clamp Meter
1. AC Voltage
2. AC Current
2b. AC High Current
3. DC Voltage
4. DC Current
5. Frequency
6. Resistance
7. Capacitance
8. Temperature Simulation (RTD )
1. 10 µV to 1000 V @ 50 Hz to 10 kHz
2a. 10 µA to 120 A @50 Hz to 10 kHz
2b. 10 µA to 1 kA / 2 kA / 3 kA / 6 kA @ 50 Hz
3. 10 µV to 1000 V
4. 1 µA to 100 A / 2 kA
5. 100 mHz to 10 MHz
6. 1 mΩ to 20 GΩ
7. 1 pF to 100 μF @ 1 kHz
8. RTD (All Types):  -200 ℃ to 800 ℃
1. MFC Fluke
2. Transconductance Amplifier (Fluke)
3.  DC Reference Standard (Fluke)
4. Reference Divider (Fluke)
5. Decade/Standard Resistance Box 
6bVoltmeter / Ampere meter / Frequency meter / Power Factor meter / Phase Angle meter / VFD meter / Multifunction meter 1. MFC
7Measure Milli ohm meter / Megaohm meter / Micro ohm meter / Earth Tester / Insulation Tester / CRM Meter / TWRM meter / Resistance100 µΩ to 500 GΩ (upto 1 kV)1. MFC (Fluke)
2. Decade/Standard Resistance Box  
8Measure Power Meter / Power Analyzer / Energy Meter (Class: 0.2, 0.5, 1.0) / Reference Energy Logger (Class: 0.02, 0.05 and coarser) 1. DC Power
2. AC Power/Energy
3. Power Factor:
1. 1 mW to 20 kW
2. 0.3 mW/mWh to 172.8 kW/kWh
3.  0.01 to UPF
1. Zera 3 Ph Ref Std (COM 3003 / COM 3000)
2. Zera AC Voltage / Current Source (MT551)
9Measure LCR Meter / Smart Tweezer1. Resistance (@1 kHz)
2. Capacitance (@1  kHz)
3. Inductance (@1 kHz)
1. 1 Ω to 100 kΩ
2. 1 pF to 100 μF
3. 100 µH to 10 H
1. Keysight LCR meter
2. Decade Resistance Box
3. Standard Capacitance Box
4. Standard Inductance Box
10Measure Temperature Indicator / Controller / Data Logger / Data Acquisition Switch UnitTemperature Simulation (RTD & Thermocouple)Thermocouple (All Types):  -250 ℃ to 2300 ℃
RTD (All Types):  -200 ℃ to 800 ℃
1. MFC (Fluke)
11Measure High Voltage Divider / H V ProbeAC & DC High Voltage1 kV to 100 kV @ 50 Hz
1 kV to 100 kV
Moonlight & R E – Capacitive Voltage Divider
12Measure Current Transformer / BridgeRatio Error & Phase Error1A to 3200A (Primary)
1A to 5A (Secondary)
Turn Ratio: 1 to 600 Turns
1. Eltel / Moonlight CT
2. Eltel / AITTS (Bridge)
13Measure Potential Transformer  / BridgeRatio Error & Phase Error1.1 kV to 132 kV (Primary)
110 V to 110 /√3 V (Secondary)
Turn Ratio: 1 to 3200 Turns
1. Moonlight PT
2. Eltel / AITTS (Bridge)
14Measure Tesla Meter/Gauss Meter / AC Magnet / DC Magnet1. AC Magnetic Field @ 50 Hz
2. DC Magnetic Field 
1. 0.1 Gauss to 2000 Gauss
2. 100 Gauss to 10000 Gauss
1. Tesla meter
2. EMF Tester
3. D C Reference Magnet
4. AC Electro Magnet
15Measure Oscilloscope1. Time Marker
2. DC Amplitude
3. AC Amplitude (1 kHz to 10 kHz)
4. Bandwidth  
1. 10 nsec to 5 sec
2. 1 mV to 100 V
3. 5 mV to 100 V
4. 50 kHz to 1000 MHz
1. MFC (Fluke)
16Measure Time Calibrator / Timer / Stopwatch / Hour meterTime Interval100 ms to 27 hours Time Interval Meter

INSTRUMENTS CALIBRATED

  • Multi Product Calibrators (Fluke/ 5550A/ 5522A/5500 Series and 5080A, Transmille, Wavetek 9100, Meatest, ZEAL etc).
  • Reference Multimeters (3.5, 4.5, 5.5, 6.5, 7.5, 8.5) (Fluke/ 8508A/ 8588A/ 8846A, Agilent, Keysight, Metravi, Keithley, Rigol etc).
  • LCR Meters.
  • Temperature Calibrators/ Data Loggers/ Super DAQ (Fluke, Zeal, Isotech etc).
  • Thermocouple Calibrators
  • Process Calibrators
  • Standard/Decade Resistance, Capacitance & Inductance Boxes.
  • HV Mega Ohm Boxes
  • Micro Ohm Meters
  • Megger (Insulation tester)
  • Transformer Winding Resistance Meters, Contact Resistance Meters
  • AC/DC Reference Standards
  • 1 & 3 Phase Power Meters & Energy Loggers, Accu-check (Fluke, Yokogawa, Hioki, Secure etc)
  • Power Analyzers (1 & 3- Phase)
  • Power Energy Reference standards of accuracy class 0.02%, 0.05%.
  • Energy Meter of Accuracy Class 0.2S, 0.5S, 1.0, 1.0S.
  • Frequency Counters and Signal Generators
  • RF Level Meters and RF Sensors & Attenuators
  • Timers/ Stop watches
  • Oscilloscopes
  • Gauss Meters & Magnets
  • Capacitive Voltage Dividers (CVD), HV Probes, HV Test Set (AC-DC High Voltage upto 100 kV)
  • HV Break Down Voltage Testers
  • CT & PT (1 A to 3200 A & 1.1 kV to 132 kV)
  • Primary & Secondary Injection Kits
  • High current Shunts
  • Relay test kits
  • Clamp Meters
  • Panel Meters (Volt Meter, Ampere Meter, Multifunction Meter, Frequency Meter, Power Factor Meter, VAF Meter, Watt Meters)
  1. DC Voltage (Source & Measure): 10 µV to 1000 V
  2. DC High Voltage (Source & Measure): 1 kV to 100 kV
  3. DC Current (Measure): 1 µA to 1000 A
  4. DC Current (Source): 1 µA to 2000 A
  5. AC Voltage (Source & Measure): 10 µV to 1000 V
  6. AC High Voltage (Source & Measure): 1 kV to 100 kV
  7. AC Current (Measure): 10 µA to 1000 A
  8. AC Current (Source): 9 µA to 6000 A
  9. Resistance (Measure): 100 µΩ to 20 GΩ
  10. Resistance (Source): 100 µΩ to 500 GΩ
  11. DC Power (Source & Measure): 1 mW to 20 kW
  12. AC Power/Energy (Measure): 0.3 mW/mWh to 172.8 kW/kWh
  13. AC Power (Source): 3 mW to 60 kW
  14. Power Factor: 0.01 to UPF
  15. Thermocouple (All Types): -250 ℃ to 2300 ℃
  16. RTD (All Types): -200 ℃ to 800 ℃
  17. AC Magnetic Field: 100 mGauss to 2000 Gauss
  18. DC Magnetic Field: 100 Gauss to 10000 Gauss
  19. Frequency (Source): 100 mHz to 1 GHz
  20. Frequency (Measure): 40 mHz to 6 GHz
  21. Current Transformer (Ratio Error & Phase Error): 1A to 3200A (Primary), 1A to 5A (Secondary)
  22. Potential Transformer (Ratio Error & Phase Error): 1.1 kV to 132 kV (Primary), 110 V to 110 /√3 V (Secondary)
  23. Turn Ratio: 1 to 3200 Turns
  24. AC Resistance: 1 Ω to 100 kΩ @ 1 kHz
  25. Inductance: 100 µH to 10 H @ 1 kHz
  26. AC Capacitance (Measure): 1 pF to 1 mF @ 1 kHz and 1 µF to 1 mF @ 100 Hz
  27. Time: 100 msec to 27 hour
  28. Time Period: 1 nsec to 5 sec
  29. Time Marker (Oscilloscope): 10 nsec to 5 sec
  30. DC Amplitude (Oscilloscope): 1 mV to 100 V
  31. AC Amplitude (Oscilloscope): 5 mV to 100 V
  32. Bandwidth (Oscilloscope): 50 kHz to 1000 MHz
  33. RF Power (100 kHz to 6 GHz): -42 dBm (5 mVpp) to 21 dBm (7.09 Vpp)
  34. pH Meter (by Simulation Method): 2 pH to 12 pH
  35. Conductivity Meter (by Simulation Method): 0.475 µS/cm to 475 mS/cm
  36. Oxidation and Reduction Potential Meter (ORPM) (by Simulation Method): -414 mV to +414 mV
  • 0.01 ppb GPS Controlled Rubidium Frequency Standard (Primary Level)
  • 0.3 ppm DC Voltage Reference Standard (Primary Level)
  • 15 ppm AC Measurement System (Primary Level)
  • 3.5 ppm High Performance Multi-Function Calibrator
  • 3.5 ppm High Performance Reference Multimeter
  • 100 ppm Power Energy Reference Standard
  • Standard CT of Accuracy Class 0.02, 0.05, 0.1, 0.2, 0.5, 1.0 up to 3200 Amp
  • Standard VT of Accuracy Class 0.2, 0.5, 1.0 up to 132 kV
CONTACT US FOR SERVICES